產品列表PRODUCTS LIST

首頁 > 技術與支持 > ABB 氧化鋯分析儀 AZ25 AZ20 產品簡介
ABB 氧化鋯分析儀 AZ25 AZ20 產品簡介
點擊次數:4812 更新時間:2017-10-20

Introduction

The Endura AZ20 is the latest in a long line of high-quality, combustion gas analyzers from ABB.

The sensor, based on a zirconium oxide cell, is mounted at the tip of the probe that is inserted in the flue duct. The resulting direct, in situ measurement provides accurate and rapid oxygen reading for combustion control / optimization and emissions monitoring purposes.

Endura AZ20 has been designed for extended periods of maintenance-free operation. The modular design with reduced component count improves the reliability and robustness of the system and simplifies breakdown repair if it occurs.

Kits containing all the parts needed to complete on-site repairs are available from ABB, ensuring that service personnel can effect repairs quickly and efficiently at minimum cost. The Endura AZ20probe has retained an easy-access cell arrangement, similar to the proven ZFG2 probe. This ensures cell replacement can be carried

out on-site using readily available, basic hand tools, even after long periods of high temperature operation.

The Endura AZ20 oxygen probe measures oxygen concentration in flue gas using an in situ 'wet analysis' method. The 'wet analysis' method avoids measurement error (typically20% of reading higher than the actual value) that is introduced by a sampling system using the 'dry analysis' method.

For customers with an existing installed ZFG2 probe with a ZMT or ZDT transmitter, a 55 V heater version of the Endura AZ20probe (the Endura AZ20 / ZFG2 replacement probe) is available.

System equipment comprises a (flue-mounted) Endura AZ20probe controlled by an integral or remote transmitter. During operation, a zirconia cell within the probe is held at a constant temperature of 700°C (1292 °F) by a probe heater and control

thermocouple assembly. If the heater control circuitry or software fails (unsafe), the probe heater power supply is switched off, ensuring the system fails 'safe' and protects the heater from an over-temp failure.

An output generated at the zirconia cell is processed in the transmitter giving a locally displayed O2 reading and a 4 to 20mA retransmission signal over any range between 0 % and 100% O2.

Optional automatic calibration (AutoCal*) enables automatic,semi-automatic or manual calibration to be performed using a probe-mounted gas control manifold mounted within the probehead. Calibration sequencing is software-controlled from thetransmitter.

Optional restrictors (where fitted*) in the probe head control thetest gas and reference air flow without the need for flowmeters.The probe requires only preset test gas and reference airpressures of 1 bar (15 psi) to maintain a constant flow of 2.2 l /min (4.662 scfh). This flow is not affected by changes in themeasured gas pressure of ± 0.35 bar (5 psi). 

Introduction

The Endura AZ25 combustion oxygen analyzer is the latest ina long line of high-quality, combustion gas analyzers fromABB. The AZ25 is designed to measure oxygen concentrationin high temperature furnace atmospheres.

The sensor, based on a zirconium oxide cell, is mounted at thetip of the probe that is inserted directly into the furnace. Theresulting direct, in-situ measurement provides an accurate andrapid oxygen reading, for combustion control / optimizationand process quality control purposes.

The in-situ technique provides a true oxygen measurementinside the process. It eliminates 'equilibrium shift' that iscommon to systems in which gas samples are extracted andcooled before measurement.

Reference air (500 to 1000 ml/min [1 to 2 ft3/hr]) is required foraccurate operation and can be supplied from an ABBmains-powered pump unit or a flow regulator unit.

Unique integrated automatic calibration

The Endura AZ25 automatic calibration system eliminates theneed for the expensive ancillary equipment required forautomatic calibration on traditional flue gas oxygen analyzersystems. ABB’s fully integrated, automatic calibration feature

controls the test gas sequence and detects test gasavailability, eliminating incorrect calibrations due to loss of testgas. Integrated automatic calibration secures the oxygenmeasurement, providing complete confidence in the analyzer's

performance.

Long-term operation without technician intervention minimizesoperating costs, total cost-of-ownership and optimizes systemaccuracy. A significant proportion of service visits to ananalyzer result in 'no fault found'. Endura AZ25's high-quality

manufacture, advanced diagnostics and automatic calibrationis targeted at reducing this wasted effort and cost.

Dependable

— measurement of O2 in a wide variety of processes

— virtually instantaneous response

— resistance to sulphurous reducing atmospheres

Versatile

— 600 to 1250 mm (2 to 4 ft) lengths

— choice of protective sheath materials

— integral thermocouple for automatic temperature

compensation

— high temperature operation up to 1400 °C (2552 °F)

Low cost-of-ownership

— in situ measurement

— no sampling system required

— automatic calibration functionality as standard

— optional automatic calibration hardware

Applications

— Furnace atmosphere monitoring:

— Annealing furnace

— Sintering of metals

— Ferrite sintering

— Whiteheart malleablizing

— Combustion measurement and control in:

—Soaking pits

—Process heaters

—Fluidized bed boilers

—High temperature incinerators

— All other combustion processes where the flue gastemperature is in excess of 600 °C (1112 °F)

 

Aluminous porcelain

Entry level sheath material suitable for applications withtemperatures up to 1250 °C (2282 °F) – limited resistant tocorrosive chemicals and mechanical and thermal stress.

Recrystallized alumina

Suitable for applications with temperatures up to 1400 °C(2552 °F) resistant to corrosive chemicals and more severemechanical and thermal stress are present.

Incoloy 800

Suitable for applications with temperatures up to 850 °C

(1562 °F) when mounted horizontally and up to 1000 °C

(1832 °F) when mounted vertically.

Offering limited resistance to corrosive chemicals and resistantmechanical and thermal stress.

愛澤工業(IZE_INDUSTRIES)銷售型號示例:

ABB 氧化鋯分析儀AZ25/.1.1.2.1.1.5.3.1.2.0.0.0.0 /STD

ABB 氧化鋯分析儀AZ25/.1.1.2.1.1.5.1.1.1.1.0.0.0.0

ABB 氧化鋯分析儀AZ25/.1.1.2.1.1.5.1.1.1.1.0.0.0.0.E./STD

ABB 氧化鋯分析儀AZ25/.1.1.2.1.1.5.3.1.2.0.0.0.0.E.Z.J/STD

ABB 氧化鋯分析儀AZ20/112112225111100C/STD

ABB 氧化鋯分析儀AZ20/1.2.2.2.1.2.2.3.3.7.1.1.2.E

ABB 氧化鋯分析儀AZ20/222122337112E

ABB 氧化鋯分析儀AZ20/2.2.2.1.2.2.3.3.7.1.1.21.0.0.E/STD

ABB 氧化鋯分析儀AZ20/222122337112E/STD

如果您有任何關于產品的相關技術咨詢或者需要產品的相關技術資料與文檔,請隨時愛澤工業(86-21-3100 6702),或者訪問www.ize-industries。。com,我們將盡全力解決您的問題。

If you have any question, please feel free to dial 0086-21-31006702 or browse our website:www.ize-industries。。com. Thank you.

主站蜘蛛池模板: 亚洲AV无码一区二区三区在线观看| 无码福利一区二区三区| 国产免费av片在线无码免费看| 人妻精品无码一区二区三区| 无码人妻一区二区三区精品视频| 自拍中文精品无码| 亚洲日韩av无码中文| 国产无码网页在线观看| 无码一区二区三区AV免费| 亚洲VA中文字幕无码毛片| 无码国内精品久久人妻麻豆按摩| 无码国产午夜福利片在线观看| 精品乱码一区内射人妻无码| 97人妻无码一区二区精品免费| HEYZO无码综合国产精品| 成人无码区免费视频观看| 久久久久久国产精品免费无码| 中文无码AV一区二区三区| 人妻无码久久精品人妻| av无码久久久久不卡免费网站| 亚洲爆乳无码一区二区三区| 中文字幕乱妇无码AV在线| av无码aV天天aV天天爽| 国内精品久久人妻无码不卡| 亚洲成a人无码亚洲成av无码 | 黄A无码片内射无码视频| 精品亚洲AV无码一区二区| 无码专区天天躁天天躁在线| 国产AV无码专区亚洲AV手机麻豆 | 亚洲精品无码少妇30P| 亚洲A∨无码一区二区三区| 免费无码H肉动漫在线观看麻豆| 性色AV无码中文AV有码VR| 无码日本电影一区二区网站| 无码人妻精品一区二区三区蜜桃| 777爽死你无码免费看一二区 | 在人线av无码免费高潮喷水| 18禁网站免费无遮挡无码中文| 亚洲精品无码专区在线播放| 无码丰满熟妇juliaann与黑人 | 国产精品无码素人福利|